Aluminum electrolytic capacitors must be measured under the correct conditions. Proper use of test fixtures and accurate instrument calibration are essential for obtaining reliable values, which are necessary for application calculations and product selection.
When measuring aluminum electrolytic capacitors, observe the following precautions:
1:Set the appropriate test parameters

Set the measurement item, test frequency, and test level according to the specifications in the product manual. Always discharge the capacitor before testing; otherwise, the test equipment may be damaged.
2:Monitor the ambient temperature

Maintain the room temperature at approximately 20–25°C, in accordance with the manual. Temperature variations can cause deviations in measurement results because aluminum electrolytic capacitors have relatively poor temperature characteristics. At low temperatures, capacitance decreases, while dissipation factor and ESR increase, affecting the accuracy of the data.
3:Calibrate the digital LCR meter correctly

Before measurement, thoroughly clean the test fixture. Then perform both short-circuit and open-circuit calibration. This enables the instrument to compensate reliably for the fixture’s resistance and measurement errors, ensuring accurate results.
4:Use and maintain the test fixture correctly

When testing a capacitor, position the fixture contacts as close as possible to the capacitor body. Correct setup and operation of the digital LCR meter allow the relevant parameters of aluminum electrolytic capacitors to be measured effectively.
Operating Principle of a Capacitance Meter

For many people, the operating principle of a capacitance meter may seem mysterious. In fact, the basic principle of an auto-balancing bridge circuit is not difficult to understand.
4TP Measurement Method
The most commonly used measurement technique for auto-balancing bridge capacitance meters is the four-terminal-pair (4TP) method. In this configuration, the HC and HP terminals are shorted together, while the LC and LP terminals are shorted together, as illustrated below.

Figure above — Four-terminal-pair (4TP) measurement method, including parasitic components
The HP and HC terminals are normally used as the CMH (capacitance-meter high) connection, while the LP and LC terminals are used as the CML (capacitance-meter low) connection.
Residual inductance and resistance are present in the measurement path, including the test cables. Stray capacitance also exists between the cables and between the device under test (DUT) and ground, as shown in the figure above. Users must perform measurement-path or cable compensation to eliminate the effects of these parasitic components. Otherwise, measurement accuracy may be significantly reduced.